Products / XES
XEScope

Mid-energy XES

XEScope

Mid-energy X-ray emission spectroscopy for chemical-state and electronic-structure research.

Energy range
1.1–8 keV
Energy resolution
Up to 0.5 eV; confirm measurement conditions

TECHNICAL SPECIFICATIONS

Specifications and configuration

Energy range
1.1–8 keV
Energy resolution
Up to 0.5 eV; confirm measurement conditions
Measurement
X-ray fluorescence emission spectroscopy

Website energy range updated to 1.1–8 keV. The existing Chinese brochure predates this update.

LABORATORY PLANNING

Prepare your laboratory

Access and load
Door ≥120 × 150 cm; floor loading 500 kg/m²; freight elevator ≥1 tonne, cabin ≥120 × 150 × 130 cm.
Power
Instrument 4 kW, chiller 5 kW, computer 1 kW; three 10 A outlets.
Utilities
50 L water, 40 L helium and an 80 × 40 cm auxiliary table.

Requirements from the XEScope installation document; confirm the electrical connection plan for the selected system.

Installation checklist (Chinese DOCX) ↓
Original installation layout · Chinese

RESOURCES & SUPPORT

Explore XEScope

Match the system to your sample, measurement target and laboratory.

Discuss a systemSample evaluation
Enlarged source figure